Coupled Markers (Na Mode) - Keysight Technologies FieldFox A Series User Manual

Table of Contents

Advertisement

How to display the Marker table
— Press Marker.
— Then
More
— Then
Marker Table On

Coupled Markers (NA Mode)

I/Q Analyzer (IQA) mode has coupled markers. but the behavior is different
than the NA mode Coupled Markers behavior. Refer to
(Waveform Only)" on page
Coupled Markers, available in NA mode only
markers that have the same number for all traces. For example, when marker 2 is created on a
trace, a marker 2 is created for ALL traces. When any marker 2 is moved, they all move together to
the same X-axis position.
— Coupled markers are allowed only in NA mode and are relevant only in multi-trace
configurations.
— You can have markers that are coupled and others that are uncoupled.
— Peak functions move the marker to the peak on the active trace, while all other markers move to
the same X-axis position.
Figure 28-3
Coupled Markers move on ALL traces at the same time (NA shown).
1. IQA mode has coupled markers, but the behavior is different than NA mode's coupled markers. Refer to
"Coupled Markers (Waveform Only)" on page
Keysight N9927-90001 User's Guide
Data Analysis Features
All about Markers
Off.
1
and allows you to simultaneously create and move
278.
278.
"Coupled Markers
601