SA (Spectrum Analyzer) Mode
How to Set Up EMF Settings (Requires EMF (Option 358), GPS (Option 307), and SA
Mode (Option 233–Mixed Analyzers))
Keysight N9927-90001 User's Guide
—
Default
—
Src/Rcvr Split
and source side antenna measurements.
—
Triaxial XYZ
when a compatible triaxial antenna's factors have
been imported) for measuring channel
measurements and EMF.
— Then Back to exit to the previous softkey menu.
e. Optional:
Verifying antenna and cable corrections are applied:
i. Press Scale/Amptd >
ON
Verify the Apply Corrections softkey is set to ON (This is the
default setting when a compatible USB antenna's correction
factors have been imported, via the System softkeys).
Verify the
X axis
are all ON. These can each be enabled or disabled
individually, but for EMF measurements all three axes should
be enabled (i.e., Antenna = Off, the applications are setting
the X/Y/Z).
Verify
Cable
where an additional cable has not been connected to the
triaxial antenna. Before a cable can be used with the EMF
measurements it needs to have corrections loaded. If you are
going to use a cable, then choose as needed:
—
Cable Off
Receiver side only antenna measurements.
to work with non-triaxial receiver
(default, when Channel Power is on and
Corrections
Antenna,
Y axis
Antenna, and
is ON. This can be disabled for applications
to open the cable correction softkeys menu
—
Edit Cable
displays a cable corrections table
that you can edit
—
New
displays a cable corrections table that
you can edit with new cable correction
content.
——Choose
Yes
cable data
——Choose
No
edit table view
—
Save Cable
Opens a menu to name and save
the cable corrections data
—
Recall Cable
Recalls the selected cable
correction file from the \Cables folder.
>
Apply Corrections OFF
Z axis Antenna
to overwrite the current
to exit the cable data
211