Calibration for NA, CAT, and VVM Modes
of the measurement WITHIN the frequency range of the waveguide Cal Kit.
Otherwise, an error message will appear during the 'Calculating Steps' portion
of the calibration.
Waveguide Cal Kits
Keysight sells two waveguide Cal Kit series: the premium 11644A series and
the economy N9911X series. Both are available online at
Effective Velocity Factor
Velocity factor is the speed at which an electromagnetic signal passes through
the transmission medium relative to the speed of light. This value is important
when distance is being calculated in DTF measurements (CAT mode) and Time
Domain (NA mode).
When the media is waveguide, the velocity factor changes with frequency.
FieldFox calculates this 'effective' velocity factor automatically. However, the
settings are different for CAT mode and NA mode.
CAT Mode - How to make Waveguide settings
These settings are necessary ONLY when making DTF measurements.
— Press Measure 1 then
— Press Meas Setup 4
— Then
— Set Media = Waveguide. Frequency Mode = BandPass is automatically
— Scroll down to Waveguide Definitions. Select the Waveguide Standard
— Press Done, then press Back.
— Cable Correction = Auto is the default setting. The Effective Velocity Factor
NA Mode - How to make Waveguide settings
These settings are necessary ONLY when your measurement requires electrical
delay or port extensions, or if using Time Domain Transform.
138
Settings
(Learn how to use the Quick Settings table in
and change Quick Settings" on page
selected for you.
being used. If your waveguide standard is NOT listed:
— Select User Waveguide. Then press Done.
— Then
DTF Cable
Edit
Cable.
— Scroll to set Waveguide Definition. The default setting is VF Corr =
Auto.
— Set the Min, Max, and Cutoff Frequencies.
is calculated automatically based on the frequencies of the waveguide
standard. To override this setting, set Cable Corr = Man.
Distance to Fault (dB)
48.)
Specifications, then
Edit/Save/Recall Cables,
Keysight N9927-90001 User's Guide
www.Keysight.com
to select a DTF measurement.
"How to view
then