Appendix B: Time-out Settings

The following procedure provides the instructions on how to set the time-out
value in the Keysight IO Libraries.
1 To launch your Keysight Connection Expert, go to Start > All Programs >
2 On the Instrument I/O on this PC panel, select your device in the USB
3 Select Send Commands To This Instrument and the Keysight Interactive IO
4 Go to Interactive IO > Options and the Options dialog box will appear.
5 Configure the time-out value based on the table provided below.
Table 4-2
Time-out settings
Points
1 to 10
11 of 50
51 to 100
101 to 500
501 to 1000
1001 to 4096
– Expected time, t = Points
NOTE
– The Points is set using the SENSe:SWEep:POINts command and the Tint is set
68
Keysight IO Libraries Suite > Keysight Connection Expert.
expanded list and right-click.
window will appear.
using the SENSe:SWEep:TINTerval command.
Tint (ms)
1 to 100
101 to 32767
1 to 50
51 to 32767
1 to 10
11 to 32767
1 to 5
6 to 500
501 to 32767
1 to 500
501 to 32767
1 to 500
501 to 32767
×
Tint
Keysight U2722A/U2723A User's Guide
Time-out range (ms)
Std IO 5000
Assign 110% of t
Std IO 5000
Assign 120% of t
Std IO 5000
Assign 120% of t
Std IO 5000
Assign 130% of t
Assign 110% of t
Assign 310% of t
Assign 120% of t
Assign 310% of t
Assign 110% of t