Agilent 8753ES System Verification and Performance Tests

Use this section of the manual if you are performing system verification or performance
tests on an 8753ES analyzer. If your instrument is an 8753ET, turn to
System Verification and Performance Tests" on page
System Verification Cycle and Kit Recertification
Agilent Technologies recommends that you verify your network analyzer measurement
system every six months. Agilent Technologies also suggests that you get your verification
kit recertified annually. Refer to the HP/Agilent 85029B 7-mm Verification Kit Operating
and Service Manual for more information.
NOTE
The system verification procedures can also apply to analyzers with
Option 075 (75 ohm analyzers) if minimum loss pads and type-N (m) to
APC-7 adapters are used.
Check to see how the verification kit floppy disk is labeled:
• If your verification disk is labeled HP/Agilent 8753D, HP/Agilent 8753E,
HP/Agilent 8753ES, HP/Agilent 8753ET Verification Data Disk, you may
proceed with the system verification.
• If your verification disk is not labeled as indicated above, you may send your
HP/Agilent 85029B 7-mm verification kit to the nearest service center for
recertification, which includes a data disk that you can use with the 8753ES.
8753ES System Verification
This system verification consists of three separate procedures:
1.
"Initialization" on page 2-6
2.
"Measurement Calibration" on page 2-8
3.
"Device Verification" on page 2-9
Analyzer warm-up time: 30 minutes
Chapter 2
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
2-87.