3.4.4 Monitoring of respiratory parameter

Based on the ventilatory parameters set by the User and on the patient's
characteristics, the system is able to monitor and measure a series of values
necessary for the patient's clinical evaluation.
At the top of the monitor, there is a light bar indicator (with scale from 0 to 80 cmH
the pressure inside the airways during the respiratory phase, in real time. The measured and
monitored values (right side of the monitor) are updated after each breath of the patient.
3-28
Tr. E (%)
Percentage of the inhaled flow with regard to the maximum peak where the
inspiratory phase ends and the expiratory phase begins.
Tr. I (L/min) (cmH2O)
Flow level (pressure) for detecting the patient spontaneous breathing.
Vte (ml)
Expired tidal volume guaranteed for the patient.
Vti (ml)
Inspired tidal volume guaranteed for each breath.
2
Press the icon under the respiratory
parameters column repeatedly to see
(in sequence) the all available
respiratory parameters.
User manual, DU33ND101