transfer functions for output power, as follows:
Pout = Vdc * (100% rated output power) / 10 VDC), with Vdc in volts, or
Pout = Vdc * (100% rated output power) / FSC VDC), with Vdc in volts.
The Full-Scale voltage value can be modified to any voltage between 2V to
10V from below SCPI command. For example, To set 5V as full scale for Power
programming send below command
Ex - SOUR:POW:PROG:FSC 5

Figure 8-8 Remote Power Programming Using 0-10 VDC Source

Remote Programming BY 4-20mA SOURCE

8.3.3
Remote Analog reference source of the Output Voltage, Current and Power can be
selected as 4-20 mA by sending the following SCPI Command:
SOUR:ANALOG:PROG:SOUR MA420
8.3.3.1 R
C
EMOTE
URRENT
Analog Reference source is selected as 4-20mA and programming mode of Output
Current as external by sending following SCPI command
SOUR:ANALOG:PROG:SOUR MA420
SOUR:CURR:PROG EXT
A 4-20mA current source is connected between Pin 19 (IPRG_VSOUR) and the return
M330461-01 Rev A
P
4-20
ROGRAMMING BY
External User Control Signal Connector
PPRG_VSOUR
33
18
A S
M
OURCE
0-10VDC
VOLTAGE
Source