OVER-Voltage programming – 10V range calibration
10.6.13
STEP
Issue a *CLS command.
1.
2.
Issue a *RST command.
Connect zero volt (0 +/- 0.003V) across PIN 5 and PIN 18 of Analog
3.
Programming connector.
Select the over-voltage Analog programming range as 10V. Also select
4.
programming mode as internal during calibration procedure.
SOUR:VOLT:PROT:PROG INT
SOUR:VOLT:PROT:PROG:FSC 10
SOUR:ANALOG:PROG:SOUR VOLT
5.
Calibrate the Offset point.
CAL:OUTP:OVERVOLT:EXTV:OFFS 0
Set the DC voltage at PIN 5 to 10V +/-0.003V. Calibrate the full scale point.
6.
CAL:OUTP:OVERVOLT:EXTV:FSC <UUT's Full-scale volt>
Unlock and store the Calibration.
7.
CAL:UNLOCK "6867"
CAL:STORE
CAL:LOCK
OVER-Voltage programming – 5V range calibration
10.6.14
STEP
1.
Issue a *CLS command.
Issue a *RST command.
2.
3.
Connect zero volt (0 +/- 0.003V) across PIN 5 and PIN 18 of Analog
Programming connector.
4.
Select the over-voltage Analog programming range as 5V. Also select
programming mode as internal during calibration procedure.
SOUR:VOLT:PROT:PROG INT
SOUR:VOLT:PROT:PROG:FSC 5
Calibrate the Offset point.
5.
M330461-01 Rev A
DESCRIPTION
DESCRIPTION