Voltage Programming - Current Source (Resistance Programming) Calibration - Ametek ASTDC60D1C-E00 Programming Manual

Rs232, usb, ethernet and ieee 488.2
Table of Contents

Advertisement

SCPI Status Implementation
SOUR:ANALOG:PROG:SOUR VOLT
SOUR:VOLT:PROG INT
SOUR:VOLT:PROG:FSC 5
5.
Calibrate the Offset point.
CAL:OUTP:VOLT:EXTV:OFFS 0
Set the DC voltage at PIN4 to 5V +/-0.003V. Calibrate the full scale point.
6.
CAL:OUTP:VOLT:EXTV:FSC <UUT's Full-scale volt>
Unlock and store the Calibration.
7.
CAL:UNLOCK "6867"
CAL:STORE
CAL:LOCK
Voltage programming – current source (Resistance
10.6.3
Programming) calibration
STEP
DESCRIPTION
Issue a *CLS command.
1.
Issue a *RST command.
2.
3.
Connect zero ohm across PIN 3 and PIN 18 of Analog Programming
connector.
Select the Analog programming source as Current source. Also select
4.
programming mode as internal during calibration procedure.
SOUR:ANALOG:PROG:SOUR CURR
SOUR:VOLT:PROG INT
5.
Calibrate the Offset point.
CAL:OUTP:VOLT:EXTI:OFFS 0
Connect 5k Ohm across PIN 3 and PIN 18 of Analog Programming
6.
Connector. Calibrate the full scale point.
CAL:OUTP:VOLT:EXTI:FSC <UUT's Full-scale volt>
Unlock and store the Calibration.
7.
CAL:UNLOCK "6867"
CAL:STORE
CAL:LOCK
10-52
Asterion DC Series Programming
M330461-01 Rev A

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sorensen asterion dc seriesAstdc125d1c-e00

Table of Contents