CAL:OUTP:VOLT:DAC 45000
16.
Let the output settle and measure the voltage with a high precision
voltmeter; this is value 4.
17. Enter the actual output voltage value of the fourth calibration point:
CAL:OUTP:VOLT:FIVEPOINT 4 <value 4>
18. Program the output of the fifth calibration point by sending the following
command string from the computer:
CAL:OUTP:VOLT:DAC 62250
19.
Let the output settle and measure the voltage with a high precision
voltmeter; this is value 5.
20. Enter the actual output voltage value of the fifth calibration point:
CAL:OUTP:VOLT:FIVEPOINT 5 <value 5>
21. To review entered data for five–point voltage calibration, issue the following
query:
CAL:OUTP:VOLT:FIVEPOINT?
22.
Send the command to calculate the calibration coefficients by the
processor.
CAL:OUTP:VOLT:CALC
23. Program the supply to unlock the non-volatile memory for calibration value
storage:
CAL:UNLOCK "6867"
24. Program the supply to store the calibration values in non-volatile memory:
CAL:STORE
Allow 15 seconds for the non-volatile memory to be updated with the
new calibration values.
25. Program the supply to lock the non-volatile memory for calibration value
protection.
CAL:LOCK
M330461-01 Rev A