Ametek ASTDC60D1C-E00 Programming Manual page 142

Rs232, usb, ethernet and ieee 488.2
Table of Contents

Advertisement

SCPI Status Implementation
CAL:OUTP:OVERVOLT:EXTV:OFFS 0
6.
Set the DC voltage at PIN 5 to 5V +/-0.003V. Calibrate the full scale point.
CAL:OUTP:OVERVOLT:EXTV:FSC <UUT's Full-scale volt>
Unlock and store the Calibration.
7.
CAL:UNLOCK "6867"
CAL:STORE
CAL:LOCK
REMOTE MONITOR OUTPUTS CALIBRATION
10.7
Remote Monitor output signals can be calibrated with set of commands as
described in below section. For detailed Pin description of the Remote
Monitor Outputs and connection diagram, Refer Section 8.4
Voltage Monitor Output Calibration
10.7.1
STEP
DESCRIPTION
Issue a *CLS command.
1.
2.
Issue a *RST command.
Connect the DMM at PIN 6 and PIN 22 to measure voltage monitor pin
3.
voltage. Set the load for approximately 25% of rated output current. Type
the following commands.
SOUR:VOLT 0
SOUR:CURR <UUT's Full-scale current>
OUTP:STAT 1
CAL:INIT:VOLT:MON
4.
Calibrate the Offset point.
CAL:OUTP:VOLT:MON:OFFS <voltage at PIN6 read from DMM>
5.
Set the output voltage to 80% of UUT's Full-scale rated voltage.
SOUR:VOLT <0.8 times UUT's Full-scale voltage>
CAL:OUTP:VOLT:MON:FSC <voltage at PIN6 read from DMM>
6.
Unlock and store the Calibration.
CAL:UNLOCK "6867"
CAL:STORE
10-60
Asterion DC Series Programming
M330461-01 Rev A