Asterion DC Series Programming

OUTPUT SCPI Command Reference

9.7.2
The letter "C" in the "SCPI" column means that the command syntax is SCPI compliant;
an "N" in the "SCPI" column means that the command syntax is not part of the SCPI
definition.
Command
OUTPut:PROTection:DELay <NRf>
OUTPut:PROTection:DELay?
OUTPut:PROTection:FOLD <0|1|2>
OUTPut:PROTection:FOLD?
OUTPut:PROTection:SENSe <boolean>
OUTPut:PROTection:SENSe?
OUTPut:PROTection:SENSe:DEFault
<Boolean>
OUTPut:PROTection:SENSe:DEFault?
OUTPut:STATe <boolean>
OUTPut:STATe?
OUTPut:TRIPped?
OUTPut:AUX
M330461-01 Rev A
External User Control Signal Connector
Description
Sets the programmable time delay executed by the
supply before reporting output protection conditions
after a new output voltage or current is specified.
Functional granularity of +/- 0.5 seconds
Returns the time delay to be executed by the supply.
Sets the foldback (program down) mode of the
supply. Valid arguments are
0 (OFF or do nothing, do not program down to zero),
1 (program down to zero upon entering constant-
voltage mode or constant power mode), or
2 (program down to zero upon entering constant-
current mode or constant power mode).
3 (program down to zero upon entering constant-
current mode or constant-voltage mode)
Returns the set foldback (program down) mode of
the supply.
0 = OFF; will not program down.
1 = will program down to zero upon entering
constant-voltage mode or constant power
mode.
2 = will program down to zero upon entering constant
current mode or constant power mode
3 = program down to zero upon entering constant-
current mode or constant-voltage mode
Sets the output voltage sense signal setting. Valid
arguments are 1/REMOTE or 0/LOCAL. When
REMOTE option is selected, voltage sense signal
must be connected at RVS connector at the rear side
of power supply.
Returns the setting of the output voltage sense
signal.
Sets the Power ON setting for output voltage sense
signal LOCAL or REMOTE. Valid arguments are
1/REMOTE or 0/LOCAL.
Returns the Power ON setting of the output voltage
sense signal.
Sets the output to zero or the programmed value;
opens or closes the isolation relay. Valid arguments
are 1/ON or 0/OFF. *RST state value is ON.
Returns the state of the output:
1 = ON 0 = OFF
Returns the integer value 1 (TRIPPED) or 0
(UNTRIPPED) state of the output.
Subsystem for controlling Auxiliary supply
SCPI
N
N
N
N
N
N
N
N
C
C
N
N
9-23