Ametek ASTDC60D1C-E00 Programming Manual page 131

Rs232, usb, ethernet and ieee 488.2
Table of Contents

Advertisement

Asterion DC Series Programming
Program the output of the first calibration point by sending the following
5.
command string from the computer:
CAL:OUTP:CURR:DAC 3275
Let the output settle and measure the output current with the current shunt
6.
and the high precision voltmeter; this is value 1.
Enter the actual output current of the first calibration point:
7.
CAL:OUTP:CURR:FIVEPOINT 1 <value 1>
Program the output of the second calibration point by sending the following
8.
command string from the computer:
CAL:OUTP:CURR:DAC 19000
Let the output settle and measure the output current with the current shunt
9.
and the high precision voltmeter; this is value 2.
Enter the actual output current of the second calibration point:
10.
CAL:OUTP:CURR:FIVEPOINT 2 <value 2>
Program the output of the third calibration point by sending the following
11.
command string from the computer:
CAL:OUTP:CURR:DAC 32000
12.
Let the output settle and measure the output current with the current shunt
and the high precision voltmeter; this is value 3.
Enter the actual output current of the third calibration point:
13.
CAL:OUTP:CURR:FIVEPOINT 3 <value 3>
14.
Program the output of the fourth calibration point by sending the following
command string from the computer:
CAL:OUTP:CURR:DAC 45000
Let the output settle and measure the output current with the current shunt
15.
and the high precision voltmeter; this is value 4.
16.
Enter the actual output current of the fourth calibration point:
CAL:OUTP:CURR:FIVEPOINT 4 <value 4>
Program the output of the fifth calibration point by sending the following
17.
command string from the computer:
CAL:OUTP:CURR:DAC 62250
M330461-01 Rev A
External User Control Signal Connector
10-49