Asterion DC Series Programming
Ramp V Example 2
9.14.5
Program a unit with no load at the output, to ramp its output voltage from 5 VDC to
25 VDC in 30 seconds upon the trigger command.
// Use SYST:ERR? after each command to verify no programming errors.
// turn on the unit.
*CLS
*RST
SOUR:CURR 33.0
SOUR:VOLT 5.0
SOUR:VOLT:RAMP:TRIG 25.0 30.0 // program voltage to ramp from the present
TRIG:RAMP
TRIG:ABORT
Power On INIT Example
9.14.6
Program a unit to power-on and initialize to 2 VDC @ 1A with an overvoltage protection
level of 3 VDC. Verify proper power-on initialization.
// Use SYST:ERR? after each command to verify no programming errors.
// turn on the unit.
*CLS
*RST
CAL:INIT:CURR 1.0
CAL:INIT:CURR?
CAL:INIT:VOLT 2.0
CAL:INIT:VOLT?
CAL:INIT:VOLT:PROT 3.0
CAL:INIT:VOLT:PROT?
// cycle power to the unit.
// note voltage is initialized to 2.0 VDC via front panel.
SOUR:CURR?
SOUR:VOLT?
SOUR:VOLT:PROT?
M330461-01 Rev A
// clear the unit to its power-on default settings.
// reset the unit.
// program output current to 33.0 A.
// program output voltage to 5.0 VDC.
// value (5.0 VDC) to 25.0 VDC in 30 secs.
// upon the trigger command.
// start ramp execution.
// turn off trigger mode.
// clear the unit to its power on default settings.
// reset the unit.
// set power-on initial current to 1.0 A.
// confirm power-on initial current setting.
// set power-on initial voltage to 2.0 VDC.
// confirm power-on initial voltage setting.
// set power-on initial overvoltage protection to 3.0 VDC.
// confirm power-on initial overvoltage protection setting.
// confirm power-on initial current setting.
// confirm power-on initial voltage setting.
// confirm power-on initial overvoltage protection setting.