Power Programming - 4-20Ma Source Calibration - Ametek ASTDC60D1C-E00 Programming Manual

Rs232, usb, ethernet and ieee 488.2
Table of Contents

Advertisement

SCPI Status Implementation
Calibrate the Offset point.
5.
CAL:OUTP:POW:EXTI:OFFS 0
6.
Connect 5k Ohm across PIN 34 and PIN 18 of Analog Programming
Connector. Calibrate the full scale point.
CAL:OUTP:POW:EXTI:FSC <UUT's Full-scale volt>
Unlock and store the Calibration.
7.
CAL:UNLOCK "6867"
CAL:STORE
CAL:LOCK
POWER programming – 4-20mA source calibration
10.6.12
STEP
1.
Issue a *CLS command.
Issue a *RST command.
2.
3.
Connect 4mA source OR 2VDC across PIN 33 and PIN 18 of Analog
Programming connector.
Select the Analog programming source as 4-20mA source. Also select
4.
programming mode as internal during calibration procedure.
SOUR:ANALOG:PROG:SOUR MA420
SOUR:POW:PROG INT
5.
Calibrate the Offset point.
CAL:OUTP:POW:EXT420MA:OFFS 0
6.
Connect 20mA source OR 10VDC across PIN 33 and PIN 18 of Analog
Programming Connector. Calibrate the full-scale point.
CAL:OUTP:POW:EXT420MA:FSC <UUT's Full-scale volt>
Unlock and store the Calibration.
7.
CAL:UNLOCK "6867"
CAL:STORE
CAL:LOCK
10-58
DESCRIPTION
Asterion DC Series Programming
M330461-01 Rev A