SCPI Status Implementation
Calibrate the Offset point.
5.
CAL:OUTP:CURR:EXTV:OFFS 0
Set the DC voltage at PIN 19 to 10V +/-0.003V. Calibrate the full scale
6.
point.
CAL:OUTP:CURR:EXTV:FSC <UUT's Full-scale volt>
7.
Unlock and store the Calibration.
CAL:UNLOCK "6867"
CAL:STORE
CAL:LOCK
Current programming – 5V range voltage source
10.6.6
calibration
STEP
Issue a *CLS command.
1.
2.
Issue a *RST command.
Connect zero volt (0 +/- 0.003V) across PIN 19 and PIN 18 of Analog
3.
Programming connector.
4.
Select the Analog programming source as Voltage source and 5V range.
Also select programming mode as internal during calibration procedure.
SOUR:ANALOG:PROG:SOUR VOLT
SOUR:CURR:PROG INT
SOUR:CURR:PROG:FSC 5
Calibrate the Offset point.
5.
CAL:OUTP:CURR:EXTV:OFFS 0
6.
Set the DC voltage at PIN 19 to 5V +/-0.003V. Calibrate the full scale point.
CAL:OUTP:CURR:EXTV:FSC <UUT's Full-scale volt>
7.
Unlock and store the Calibration.
CAL:UNLOCK "6867"
CAL:STORE
CAL:LOCK
10-54
DESCRIPTION
Asterion DC Series Programming
M330461-01 Rev A