REMOTE ANALOG PROGRAMMING AND
EXTERNAL USER CONTROL INTERFACE
INTRODUCTION
8.1
This section provides illustrations of the Asterion DC power supply's rear panel layout,
which differs among the Asterion DC models. For example, Refer to Figure 8-1.
Regardless of the layout, the component functions are common across all models and
those that are pertinent to the Remote Analog Programming and External user control
Interface Connector is described here. The analog programming and external user
interface control functionality can be exercised from either of the communication
interfaces (RS232, USB, Ethernet and IEEE 488.2 GPIB).
1. Remote Analog Programming and External User Control Interface Connector
REMOTE ANALOG PROGRAMMING
8.2
CONNECTOR DETAILS
A 44 -pin Subminiature-D connector (Refer to Figure 8-2) located at the rear panel
provides remote analog programming and external user interface control signals to
increase the user's operating control of the supply. The mating receptacle is
NORCOMP 180-044-102L001 with 44 Pin male terminals. The connector terminals
accommodate wire sizes from #24 - #30.
M330461-01 Rev A
Figure 8-1. Rear Panel - Analog Programming Connector
CONNECTOR
8
8-1