SCPI Status Implementation
CAL:OUTP:CURR:DAC 65535
5.
Set the output voltage sensing to "Local"
OUTP:SENS LOCAL
Program the output of the first calibration point by sending the following
6.
command string from the computer:
CAL:OUTP:VOLT:DAC 3275
7.
Let the output settle and measure the voltage with a high precision
voltmeter; this is value 1.
Enter the actual output voltage value of the first calibration point:
8.
CAL:OUTP:VOLT:FIVEPOINT 1 <value 1>
Program the output of the second calibration point by sending the following
9.
command string from the computer:
CAL:OUTP:VOLT:DAC 19000
10.
Let the output settle and measure the voltage with a high precision
voltmeter; this is value 2.
11. Enter the actual output voltage value of the second calibration point:
CAL:OUTP:VOLT:FIVEPOINT 2 <value 2>
12. Program the output of the third calibration point by sending the following
command string from the computer:
CAL:OUTP:VOLT:DAC 32000
13.
Let the output settle and measure the voltage with a high precision
voltmeter; this is value 3.
14. Enter the actual output voltage value of the third calibration point:
CAL:OUTP:VOLT:FIVEPOINT 3 <value 3>
15. Program the output of the fourth calibration point by sending the following
command string from the computer:
10-44
Asterion DC Series Programming
M330461-01 Rev A