Ametek LAND APS42 Installation Manual
Ametek LAND APS42 Installation Manual

Ametek LAND APS42 Installation Manual

Auxiliary power supply unit

Advertisement

Quick Links

APS42
INSTALLATION GUIDE
PUBLICATION N
810401
O
LANGUAGE: ENGLISH
Auxiliary Power Supply Unit, APS42
i
Q U A L I T Y C U S T O M E R S O L U T I O N S

Advertisement

Table of Contents
loading

Summary of Contents for Ametek LAND APS42

  • Page 1 APS42 INSTALLATION GUIDE PUBLICATION N 810401 LANGUAGE: ENGLISH Auxiliary Power Supply Unit, APS42 Q U A L I T Y C U S T O M E R S O L U T I O N S...
  • Page 2 IMPORTANT INFORMATION - PLEASE READ Health and Safety Information Read all of the instructions in this booklet - including all the WARNINGS and CAUTIONS - before using this product. If there is any instruction which you do not understand, DO NOT USE THE PRODUCT.
  • Page 3 Web: www.landinst.com For further details on all AMETEK Land offices, distributors and representatives, please visit our websites. Return of Damaged Goods IMPORTANT If any item has been damaged in transit, this should be reported to the carrier and to the supplier immediately.
  • Page 4 APS42 Auxiliary Power Supply Preface Installation Guide...
  • Page 5: Table Of Contents

    Auxiliary Power Supply APS42 Contents Introduction General Description Installation Requirements 1.3 Specification Installation Details Installation Dimensions Electrical Connections Maintenance Replacing the Power Supply Module Installation Guide Preface...
  • Page 6: Introduction

    INTRODUCTION...
  • Page 7: General Description

    APS42 Introduction 1.1 General Description The APS42 is an Auxiliary Power Supply unit designed to be used in conjunction with the AMETEK Land Model 4200+ range of Continuous Opacity Monitoring Systems. Model 4200+ APS42 For more information on the Model 4200+, refer to the Model 4200+ User Guide (Publication Nº...
  • Page 8: Installation Details

    INSTALLATION DETAILS...
  • Page 9: Installation Dimensions

    Auxiliary Power Supply APS42 Installation Details 2.1 Installation Dimensions Installation Guide 2 - 1...
  • Page 10: Electrical Connections

    APS42 Auxiliary Power Supply 2.2 Electrical Connections Instrument Connections Customer Connections Mains Input Earth Mains Input Live Mains Input Neutral +24V dc, Cable A, Red 0V, Cable A, Blue Cable A, Screen Transceiver Shutter A Transceiver Shutter B Retro Shutter A Retro Shutter B Transceiver Blower Pressure Switch (Common) Transceiver Blower Pressure Switch (NO)
  • Page 11: Maintenance

    MAINTENANCE...
  • Page 12 Mains input terminals 24V DC output terminals Fig. 3-1 APS42 Power Supply Module...
  • Page 13: Replacing The Power Supply Module

    Auxiliary Power Supply APS42 Maintenance 3.1 Cleaning Clean the outside surfaces of the unit with a soft, dry cloth. 3.2 Replacing the Power Supply Module (Land Part Nº 810229) The APS42 is designed to be maintenance-free. The only user-replaceable component is the internal Power Supply Module (Land Part Nº 810229). WARNING Risk of electric shock.
  • Page 14 APS42 Auxiliary Power Supply 2 - 2 Installation Guide...
  • Page 15 Gas Monitoring Products Standard Terms of Warranty Land Instruments International warrant all products of our own manufacture to be within specified limits of calibration, if any, when despatched from our works. Also that they are free from defects in material and workmanship for normal use and service when used for their intended purpose within the limits of our specification.
  • Page 16 4200+ HIGH PRECISION OPACITY AND DUST EMISSIONS COMPLIANCE MONITOR AMETEK Land China Service AMETEK Land India Service Land Instruments International AMETEK Land - Americas Dronfield, S18 1DJ, 150 Freeport Road, Pittsburgh, Part A, 1st Floor No. 460 Divyasree N R Enclave,Block A,...