Power Programming - 10V Range Voltage Source Calibration - Ametek ASTDC60D1C-E00 Programming Manual

Rs232, usb, ethernet and ieee 488.2
Table of Contents

Advertisement

SCPI Status Implementation
Calibrate the Offset point.
5.
CAL:OUTP:CURR:EXT420MA:OFFS 0
Connect 20mA source OR 10VDC across PIN 19 and PIN 18 of Analog
6.
Programming Connector. Calibrate the full scale point.
CAL:OUTP:CURR:EXT420MA:FSC <UUT's Full-scale volt>
7.
Unlock and store the Calibration.
CAL:UNLOCK "6867"
CAL:STORE
CAL:LOCK
Power programming – 10V range voltage source
10.6.9
calibration
STEP
Issue a *CLS command.
1.
2.
Issue a *RST command.
Connect zero volt (0 +/- 0.003V) across PIN 33 and PIN 18 of Analog
3.
Programming connector.
4.
Select the Analog programming source as Voltage source and 10V range.
Also select programming mode as internal during calibration procedure.
SOUR:ANALOG:PROG:SOUR VOLT
SOUR:POW:PROG INT
SOUR:POW:PROG:FSC 10
Calibrate the Offset point.
5.
CAL:OUTP:POW:EXTV:OFFS 0
6.
Set the DC voltage at PIN 33 to 10V +/-0.003V. Calibrate the full scale
point.
CAL:OUTP:POW:EXTV:FSC <UUT's Full-scale volt>
Unlock and store the Calibration.
7.
CAL:UNLOCK "6867"
CAL:STORE
CAL:LOCK
10-56
DESCRIPTION
Asterion DC Series Programming
M330461-01 Rev A

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sorensen asterion dc seriesAstdc125d1c-e00

Table of Contents