External User Control Signal Connector
Figure 8-12 Remote Overvoltage Programming Using DC Voltage Source
REMOTE MONITOR OUTPUTS
8.4
DC Asterion provides independent analog voltage monitor signals for the Output
Voltage, Current and Power.
1. Voltage Monitor Output is in Pin 6, Return Pin 22.
2. Current Monitor Output is in Pin 21, Return Pin 22.
3. Power Monitor Output is in Pin 35, Return Pin 22.
4. Full Scale Monitor voltage would correspond to rated output of the power supply.
Default Full Scale Monitor Output is 10V. Same can be changed from 2V to 10 V
by issuing following SCPI command. For example, to set to 5V as full scale for
voltage monitoring send below command:
SOUR:VOLT:MON:FSC 5
5. For the Current monitor to change the full-scale value to 5V, following SCPI
command to be used.
SOUR:CURR:MON:FSC 5
6. Similarly, for the Power monitor to change the full-scale value to 5V, following SCPI
command to be used.
SOUR:POW:MON:FSC 5
REMOTE DIGITAL INPUT AND OUTPUT
8.5
Auxiliary power output
8.5.1
Asterion DC power supply has the option of two auxiliary power outputs 5 V and
15V. These auxiliary power outputs can be controlled from the power supply
remotely or by giving appropriate signals on the digital input enable pins provided
for the same.
8-18
Asterion DC Series Programming
OVPRG_VSOUR
5
18
M330461-01 Rev A
0-10VDC
VOLTAGE
Source