Figure 1-11 shows a typical Main Screen with patients in alarm. Note that Patient Sector
CCU2 has the alarm message ***TACHY 160>140. Patient Sectors in alarm are backlighted
in blue for easy identification, and the alarm message window is backlighted in the color of
the alarm severity. For example, the CCU2 message shows a red alarm (***) message.
Table 1-8 summarizes the various alarms that can occur, including the corresponding alarm
sound, alarm message location, ECG recording generation, and alarm meaning. Alarm
messages are described in the IntelliVue Information Center Instructions for Use.
Alarm
Red (***)
Yellow (**)
Yellow (**)
Arrhythmia
Yellow (**)
Nurse Call on
Telemetry
Hard INOP (*)
Soft INOP (*)
Arrhythmia
In addition to patient monitoring, the Information Center system provides arrhythmia
Monitoring
monitoring for both bedside and telemetry monitored patients. The multi-lead arrhythmia
monitoring algorithm is designed for adult, pediatric, and neonatal patient ECG waveforms
ECG recording is generated for preset periods preceding and following the alarm
event (Red and Yellow alarms only)
Table 1-8. Information Center Alarm Priorities
Sound
Traditional/CareNet
sound: continuous, high-
pitch rapid tone
IEC/ISO sound: repeated
bursts of five rapid high-
pitch beeps
Traditional/CareNet
sound: continuous,
medium-pitch tone
IEC/ISO sound: repeated
bursts of three rapid
low-pitch beeps
Traditional/CareNet
sound: noncontinuous,
medium pitch tone
(~10s)
IEC/ISO sound: two
rapid low-pitch beeps
Traditional/CareNet
sound: noncontinuous,
medium pitch tone (~8s)
IEC/ISO sound: two
rapid low-pitch beeps
Traditional/CareNet
sound: continuous, slow
low-pitch tone
IEC/ISO sound: repeated
bursts of two slow low-
pitch beeps
none
Message
Recording Meaning
On Display
Yes
with ***
On Display
Yes
with **
On Display
Yes
with **
On Display
Yes
with **
On Display
No
with no *
On Display
No
with no *
Introducing the IntelliVue Information Center System
Clinical Applications
Life threatening, e.g.
AYSTOLE
non-arrhythmia alarm limit
violation (except for HR,
which is an Arrhythmia
alarm)
Arrhythmia yellow alarm
detected
Nurse Call button on
telemetry transmitter
depressed and system
configured to alarm
Inoperative condition that
prevents monitoring, e.g.
LEADS OFF, or has direct
effect on the patient, e.g.
NBP CUFF OVERPRESS
Inoperative condition that
prevents signal processing,
e.g., NOISY ECG.
Monitoring continues during
this INOP condition.
1-33