System Events

To access System events that are designated Error and Warning, follow the steps for
Application Events on page 7-27, except check System in Step 2. The following table
gives entries that can appear as System events.
Source
Description (Event Code)
SDN
Error: 513, SDN RAM error
Error: 514, SDN signature RAM error
Error: 515, SDN autopoll error
Error: 516, SDN TB2 error
Error: 517, SDN chip access not in dead time at start of dead
time. [(No SYNC TAP, (i.e., no SDN connection to IC)]
Error: 519, Background not complete
Error: 769, Host not responding to interrupt from SDN card.
Error: 770, Write to host did not complete - DATA LOST.
Error 771, Read from host did not complete.
Error: 896, The card was told to read 0 bytes from the SDN
driver
Error: 897 The driver is trying to send too much data to the
card. The transfer did not complete.
Error: 898, The amount of data read by the card from the driver
does not agree with the expected count. The transfer did not
complete.
Error: 899, The transfer count indicating the amount of data
that the driver wants the card to read exceeds the amount that
can be read by the AMCC chip. The transfer did not complete.
Error: 900, The transfer count indicating the amount of data
that the driver wants the card to read is old. The transfer did
not complete.
Error: 901 The message header contained in the driver to card
data is not a valid header. Transfer did not complete.
Error: 902, The size of the recorder message from the driver to
the card is incorrect. The transfer did not complete.
Error: 903, The driver to card read appeared to be successful
but the transfer count to read did not agree with the amount
read. The transfer did not complete.
Error: 904, The recorder ID in the driver to card recorder
message is not a valid ID. Data not sent to the recorder.
Table 7-8. System Events
Corrective Action
Replace SDN Interface Card
Replace SDN Interface Card
Replace SDN Interface Card
Replace SDN Interface Card
Check SDN cable connection to the Philips PC
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reboot Philips system. If problem persists,
replace SDN Interface Card
Maintenance, Troubleshooting, and Repair