Alarm Annunciation; Priorities; Annunciation - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

Clinical Applications
available as an option for both systems, but it is not required when used with the Information
Center because the Philips system provides ST Segment analysis of telemetry data as a
standard feature.
M1403A/J Digital UHF
Telemetry System
Alarm
Information Center software also provides extensive alarm annunciation capability and

Annunciation

flexibility. Alarm events are detected by comparing physiological data against preconfigured
limit values. Alarms detected by Information Center software include:
Note
The Information Center displays alarm events that it detects as well as those generated by
networked monitors. However, the Information Center only controls alarm limits and on/off
status for alarms that it detects.

Priorities

Information Center software prioitizes alarms into 3 levels of severity:
Annunciation
Alarms are announced by the following indicators:
1-32
Introducing the IntelliVue Information Center System
Figure 1-20 Telemetry Systems for Information Center Monitoring
Arrhythmia alarms (for data from both bedside and telemetry monitors)
ST Segment alarms (for data from telemetry monitors only)
Red
*** life threatening (e.g. Asystole)
Yellow **
alarm limit exceeded (e.g. high heart rate)
Cyan
inoperative (INOP) condition (e.g. patient leads off)
Patient Sector turns blue (Red and Yellow alarms only)
Alarm Tone sounds that indicates the alarm severity (no sound for soft INOPs)
Alarm message is displayed in Patient Sector and Patient Window of the same color
as the alarm.
– For rate alarms, the message indicates:
— parameter in alarm
— maximum or minimum value of that parameter
— alarm limit that was violated (e.g. HR 134>120)
– For event alarms, the message indicates the event that caused the alarm (e.g.
Asystole)
M2600 Philips
Telemetry System