Philips M3150B Installation And Service Manual page 239

Information center system
Table of Contents

Advertisement

Field
Restore Patient ADT
Settings?
Start Restoring:
Note
After restoration is complete, the ConfigWizard must be run until all the settings have been
verified in each window. Restoration of the settings from earlier releases may not populate all
the settings that are available in release E.01.
Note
The Patient Data Transfer Bandwidth Utilization setting must be manually configured. It is
not archived. See "Patient Data Transfer - Bandwidth Utilization" on page 7-75.
Warning
Patient ADT settings should only be restored if the Archive was made very recently and
contains information on patients presently being monitored. Patients admitted after the
archive are not restored.
If patient ADT settings are restored, they should be verified to assure that patient ADT
data are accurately displayed and stored.
Note
If the message "Restore from Archive failed. Error: Error performing inpage operation"
appears, reboot the device and retry the Restore.
Procedure The procedure for restoring archived configuration settings from an Archive disk is as
follows.
Step 1. Insert the disk into the A:\ drive.
Step 2. Enter the correct configuration file name from the disk in the Archive path: field.
If you do not recall the correct file name:
Description
Select whether information on patients that were on Information Centers at the time the Archive
disk was created should be restored as admitted patients. Clicking a check in the box means that
patient names will be restored. No check means they will not be restored and no patient names
will appear on the Main Screen.
Patient ADT settings that will be restored for each patient on the Network are the following:
Patient Name
Medical Record Number #
Admit state
Paced Status
Patient Type
Initiates the Restore process once the correct archive file appears in Archive path
– Click Browse to display the File Manager files.
– Click on 3 1/2 Floppy A:\ to display the files on the disk.
– Click on the correct file name in the A:\ list
– Click OK and the file name will appear in the Archive path: field.
Software Installation & Configuration
Configuration
6-13