SDN Interface
The following table gives Symptoms, Possible Causes, and Corrective Actions for
Card
problems with the SDN Interface Card.
Symptom
"No data from bed" error
message on Main Screen.
SDN data-source malfunction
Philips system malfunction
Table 7-13. Troubleshooting the SDN Interface Card
Possible Cause
Faulty SDN cable
SDN Interface Card malfunction
Software malfunction
Faulty bedside monitor
Faulty telemetry monitor
System configuration
Corrective Action
Check all SDN cables
- Run SDN Card Diagnostic
- Check Event Log for causes
- Shut down system and reboot
- Shut down system and reboot
- Reinstall Philips application software
- Reinstall Windows Operating System
software
Troubleshoot bedside monitor
Troubleshoot telemetry monitor
An Information Center system must be on
Branch 0 of the SCC.
Maintenance, Troubleshooting, and Repair
Troubleshooting
7-115