The following table gives Symptoms, Possible Causes, and Corrective Actions for
Card
problems with the SDN Interface Card.
Symptom
"No data from bed" error
message on Main Screen.
SDN data-source malfunction
Philips system malfunction
Table 7-13. Troubleshooting the SDN Interface Card
Possible Cause
Faulty SDN cable
SDN Interface Card malfunction
Software malfunction
Faulty bedside monitor
Faulty telemetry monitor
System configuration
Corrective Action
Check all SDN cables
- Run SDN Card Diagnostic
- Check Event Log for causes
- Shut down system and reboot
- Shut down system and reboot
- Reinstall Philips application software
- Reinstall Windows Operating System
software
Troubleshoot bedside monitor
Troubleshoot telemetry monitor
An Information Center system must be on
Branch 0 of the SCC.