*System Serial Number
*Device Serial Number
Software Release
*Contact
*Optional Fields
Purchased Options displays the purchased options that are active on the device.
# of Displays
Viewable Display Width
# of ICs
# of Patients
System Serial # of the Server. It is found on the label on the top of the Server as System
Device Serial # of the Database Server. It is also found on the label on the top of the Server as
Product.
Read Only. The number of the application software revision installed -- E.xx.xx for present
software
Service resource information for users. Names and contact information (telephone numbers,
beeper numbers, etc.) of persons or contractors assigned as service support can be entered in this
field along with any other support information of benefit to users, e.g. service contract numbers.
Information is entered by keyboard and has a 255 character limit.
Entries are not required in fields preceded by an *
Note—If configuring a Dual Display system, both displays must be connected in order to
select 2 in this field.
Note—If a display is disconnected or fails on a system configured for dual display, the resting
display comes up in an unusable state. In such a case when a second display replacement is
not available, the number of displays must be changed to 1 for the system to work properly.
Number of displays for the device
• Networked Information Center: 1 (default) or 2 (Option C01)
• Database Server: disabled
• Local Database Information Center: 1 (default) or 2 (Option C01)
• Information Center Client: 1 (default) or 2 (Option C01)
• Patient Link: disabled
Sets the viewable sweep speed parameter. Use this list as a guideline in determine what value
is to be used for the display size of the device being configured.
• M3161 (medium) CRT Display: 308 mm
• M3162 (large) CRT Display: 381 mm
• 862100 Medium Flat Panel Display: 338 mm
• 862428 Medium Touch Flat Panel Display: 338 mm
• 862103 Large Flat Panel Display: 359
# of Information Centers that can be connected to the device
• Networked Information Center: not applicable
• Database Server: 2 (Option A02), 4 (Option A04), 6 (Option A06), or 8 (Option A08)
• Local Database Information Center: not applicable
• Information Center Client: not applicable
• Patient Link: not applicable
# of patients whose data can be stored by the M3154 Database Server.
• Networked Information Center: 4 (Option A04), 6 (Option A06), 8 (Option A08), 12
(Option A12), or 16 (Option A16)
• Database Server: disabled
• Local Database Information Center: 4 (Option A04), 6 (Option A06), 8 (Option A08), 12
(Option A12), or 16 (Option A16)
• Information Center Client: 4 (Option A04), 6 (Option A06), 8 (Option A08), 12 (Option
A12), or 16 (Option A16)
• Patient Link: 4 (Option A04), 6 (Option A06), 8 (Option A08), 12 (Option A12), or 16
(Option A16)