Table 7-6. Error and Status Messages on Main Screen
Message
Recorder Error Messages
XXX recorder out of paper
(XXX = Left, Center, or Right)
XXX recorder door is open
No recorder connected
XXX recorder hardware fault
Recorder Rack
Power Supply Fault
No Recorder
Printer Status Messages
Appxx print job - in progress
sent to printer
- waiting for
printer
- failed
where Appxx is the name of the
application (e.g. Trend Review)
Check printer for problems
Printer is busy
Printer needs attention
No printer connected
Other Messages
Move mouse over blue sector to silence
alarms
To Restore Normal Operation - Press
"Restart Network" in Patient Window
Operating in Local DB -- See User's
Guide for Details; Check Patient and
Unit settings
System problem -- contact service.
Some functions disabled
Patient Database Server Disk Failure --
contact service
The sound subsystem failed -- No Alert
Sound
Possible Cause(s)
Recorder is out of paper
Recorder door is open
- No Recorder is in Recorder Rack
- Recorder is in Recorder Rack
Recorder is inoperable
There is a fault in the Recorder Rack or
Recorder Rack Power Supply
-
- print job is being sent to the printer
- printer has received the job, printing
is in process
- print job is spooled and waiting for
printer to become available
- print job has failed
- printer has a paper jam
- out of paper
- printer off line
- printer is presently printing
- printer problem exists
System does not recognize any connected
printer
No mouse movement within 30 seconds
following an alarm
Connectivity to Server has been restored
after a power failure
Connectivity to Server has failed
Server has detected a failure on one of its
disks
Device was unable to generate an alarm
or inop sound
Corrective Action(s)
Replace paper roll
- Close Recorder door
- Replace Recorder
- Insert Recorder in Recorder Rack
- Reseat Recorder in Recorder Rack
- Replace Recorder or Rack
Replace Recorder.
- Reseat Recorder in Recorder Rack
- Check Recorder Rack and Power
Supply
- None
- None
- None, If problem persists, check
printer, printer spooler or hub
- Resend print job
- remove paper jam
- add paper to printer
- consult printer User's Guide
- None
- see steps above
- check printer connection
- check printer configuration
Use the mouse to silence the alarms
Press Restart Network in Patient
Window.
System reboots and monitoring will be
unavailable for about 5 minutes
Check User's Guide to review limited
operations
Check Patient and Unit settings
Check Network/Server connectivity
Replace failed disk
System will automatically reboot.
If problem persists, replace sound card