Troubleshooting

Server Connectivity

7-16
Maintenance, Troubleshooting, and Repair
Server connectivity -- are patient monitoring data being correctly stored by the
Server and accessible from the Server by Information Centers and Clients? Each
Server/Device connection should be checked to determine if stored patient
monitoring data can be reviewed.
18.5
Information Centers
18.5
Figure 7-4 Server Connectivity Investigation
The following table provides a brief guideline for identifying sources of Server
connectivity problems.
Table 7-2. Server Connectivity Troubleshooting
Assessment
Are patient data (waveforms and
parameters) being received by the Server?
Are patient data (waveforms and
parameters) being stored correctly by the
Server?
Are all Network devices operational?
Are all devices connected?
Switch
Server
18.5
Investigation
• Check that Information Centers are not in
Local Database Mode
• Check Review Applications for each
connected patient and verify that patient
data have been correctly stored
• Check Status LEDs on each Network
device -- switch, repeaters, media
translators
• Check Status Log (See Caution)
• Ping each device from the Server
• Check Link LEDs
18.5
Clients
18.5