Philips M3150B Installation And Service Manual page 315

Information center system
Table of Contents

Advertisement

Description
Source
(Event Code)
dataserver
Error: 2000
Error: 2001
Error: 2002
Error: 2004
Error: 2005
Error: 2006
Error: 2007
Error: 2008
Error: 2009
Error: 200B
Error: 200C
Error: 200D
Error: 200E, 200F,
2010, 2011, 2106,
2017, 2018, 2019
Error: 201A
Error: 201B
Error: 2021
Error: 2022
Error: 2023
Error: 2024
Table 7-7. Application Events
Explanation
Monitoring conflict or Central Station
# conflict. A bed is being monitored by
more than one Information Center
Patient Data Store told DataAcquisition
to tune more than 16 beds
simultaneously
Network error. SCC probably
disconnected or turned off.
Length of message block received from
Redirector does not equal the first word
of the buffer
Invalid link to next SDN message.
Corrupt incoming data.
SDN message too small
HRC check failed. Corrupt incoming
data
Invalid link to next SDN message.
Corrupt incoming data.
Invalid high order bits in SDN
message. Corrupt incoming data.
Virtual Terminal Protocol could not
connect to arrhythmia controls object.
Virtual Terminal Protocol could not
connect to database alarms object
Virtual Terminal Protocol error in
GetAlarmWave
Low level Data Control Protocol error
Data Acquisition Receive thread is
shutting down
Low level Data Control Protocol error
Branch number changed. User altered
SCC configuration while system was
running
SDN Interface Card reset
Data Link Control header incorrectly
terminated
SDN Data block incorrectly terminated
Corrective Action
Correct SDN branch assignments in SDN
Config and Display Setup to assure that
each monitor is assigned to only one
Central Station
Contact Response Center
- Check SCC cable and SDN interface
connection
- Check that SCC is turned on.
Contact Response Center
Contact Response Center
Contact Response Center
Contact Response Center
Contact Response Center
Contact Response Center
Contact Response Center
Contact Response Center
Contact Response Center
No action required
Not an error, no action required
No action required
- Check SCC cabling
- Shut down Information Center and restart.
System should recognize new
branch and start normally.
Shut down Information Center and restart
Reseat or replace SDN Interface Card
Reseat or replace SDN Interface Card
Maintenance, Troubleshooting, and Repair
Troubleshooting
7-29