Philips M3150B Installation And Service Manual page 265

Information center system
Table of Contents

Advertisement

Setup
Remove
Equipment Setup for Monitor Labels field is used to assign M3/M4/IntelliVue Monitor
Labels to Bed Labels.
The left column displays icons for the Monitor Labels for M3/M4/IntelliVue Patient
Monitors on the network. These are the labels setup in the Monitor Label field for each
Information Center on the network. Clicking on (highlighting) the Monitor Labels icon
causes the patient monitors for that Information Center to be displayed.
The other columns are the following:
Field
Label #
Monitor Label
Assigns monitoring Equipment on each SCC branch # to an Information Center for monitoring:
• Click on (highlight) an SCC in the left column to display its LBN #s.
• Click on (highlight) an LBN # to select it for assignment.
• Click on Setup to bring up the Map Monitor to LBN #X window (where X is the # of
the LBN selected).
Fields in the Map Monitor to LBN #X window are:
• Equipment type is for specifying the type of monitoring equipment -- Hardwired or
Telemetry -- connected to the selected SCC branch number. Clicking in the circle
preceding the monitor type selects it.
• Equipment Label is for entering an Equipment Label for telemetry monitors. Because a
telemetry monitor is not associated with a specific bed, its Equipment Label should be
different from its Bed Label, e.g. Tel1. This field is activated if Telemetry is selected, as
shown in the above right window , and disabled (grayed out) if Hardwired is selected, as
shown in the above left window. For hardwired monitors, the Equipment Label is
automatically made the same as the Bed Label.
• Setup equipment only is for assigning monitoring equipment to a branch number
without giving it a Bed Label. This allows for telemetry monitors to be available to the
user in Sector Setup but with no Bed Label or Patient Sector assignment. Clicking a
selects this option and the Located at bed: field will be disabled.
check in the box
Note—When adding new equipment (e.g. IntelliVue Patient Monitors) to an existing system in
Flex mode, there must be an additional label added for each bedside Monitor label that was
added. This additional label is required in order to enable this option.
• Located at bed: is for selecting the Bed Label to be assigned to the monitoring
equipment. Bed Labels and Unit names in the Bed Labels field appear in this window.
Clicking on (highlighting) a Bed Label in the list assigns that Bed Label to the monitoring
equipment. Up/down buttons to the right of the field permit scrolling the list up and down
– Clicking OK stores the entered information, which then appears for the LBN selected in
the list.
– Clicking Cancel closes the Map Monitor... window and returns to the Equipment
Setup window
Removes monitoring equipment from an LBN. The procedure is as follows:
• Click on (highlight) a branch number in the LBN column to select the associated equipment
for removal.
• Click Remove equipment and the monitoring equipment for the selected branch number
will be removed from the list.
Note—When removing mapped equipment, a message displays: "This will remove all
selected beds. Are you sure?" This must be confirmed before the equipment is removed.
Description
Identifies the order of the M3/M4/IntelliVue monitors in the list
Label assigned to the M3/M4/IntelliVue monitor
Software Installation & Configuration