Procedure; Setup Lbn Procedure - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

Configuration
Step 5. Select an icon -- SCC, Switch or Monitor Labels -- in the left column of the
If an SCC icon is selected, follow the Setup LBN Procedure that follows:
If a Monitor Labels icon is selected, skip to the Setup Monitor Procedure.
If a Switch icon is selected, skip to the Setup Switch Procedure.
Setup LBN
Step 1. Assign SDN monitoring Equipment (hardwired and telemetry) on SCC Logical

Procedure

Select the type of patient monitor attached to that LBN # -- Hardwired or Telemetry -- by
clicking in the circle preceding it.
Step 2. Repeat Step 6 for each LBN # and SCC in the list.
6-44
Software Installation & Configuration
Type in the desired label prefix (e.g. Bed, Tel, Mon) in the Prefix: box using the
keyboard.
Type in the first number (typically 1) of the Label sequence in the Start with Index:
box.
Type in the desired number of labels with the Prefix entered above in the Number of
labels: field. Typically this is the number of Beds, Telemetry Equipment, or M3/
M4/IntelliVue Monitors in the clinical Unit.
For Bed Labels, click on the clinical Unit where these beds reside from the
Associated with Unit field. This field is grayed out for Telemetry Equipment and
Monitor labels.
Click OK to enter these Labels in the Labels list.
Equipment Setup for ... field.
Branch Numbers (LBNs) to Bed Labels as follows:
Click on (highlight) the first SCC in the left column of the Equipment Setup for
Selected SCC field.
Click on (highlight) the first LBN # in the LBN column to activate it for assigning a
patient monitor.
Click Setup to bring up the Map Monitor to LBN #X window for that LBN #.
If Telemetry was selected, select a Label for that telemetry monitor (e.g. Tel1) in
the Equipment Label box.
If Hardwired was selected, proceed to the following step.
If this patient monitor is to be assigned a Bed Label, click on (highlight) the Bed
Label to be assigned to that patient monitor from the Located at bed list.
If this patient monitor is not to have a bed label, click a check
equipment only box.
Click OK to enter that monitor and Bed Label assignment in the Equipment Setup
for Selected SCC.
in the Setup