Test And Inspection Procedures - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

Action
Type:
Press Enter
Response if OK
Type:
Press Enter
Response if OK
Using any phone,
- dial the modem phone # to hear the modem tones
- hang up
Response if OK
Press Enter
Response if OK
Type:
Press Enter
Response if OK
Step 10. Exit the New Connection - Hyperterminal window by clicking on the X in the
Step 11. Shutdown and restart the PC or server by:
Test and
This section is intended for Philips Cardiac and Monitoring Systems Service Providers. It
Inspection
documents requirements for test, inspection, and reporting of results for Information Center
Procedures
systems to help assure safe and reliable operation.
Note
The tests and inspections in these tables must be followed by Philips Service Providers when
the Philips system is installed and after any service event.
Table 8-2 describes which tests shall be performed for Philips system components --
Information Centers, Clients, Database Servers, Clinical Network components -- for each type
of service event.
Table 8-1. Modem Test Procedure
D5000A
Model:
P2478U
P2478W
M3168A
AT&T0
000
OK
ATS0=1
OK
Ring
no carrier
AT&F1
OK
upper right of the message box and then Yes to return to the Windows Main Menu.
Click on Start
Click Shut Down
Select Restart the computer?
Click Yes
Notes
- last character is a 0 (zero)
- ends testing
- fourth character is a 0 (zero)
- auto answer on first ring
resets the modem to factory defaults
Start-up and Configuration
Configuring the System
8-11