Philips M3150B Installation And Service Manual page 263

Information center system
Table of Contents

Advertisement

An invalid character(s) has been entered in the label prefix field. Only SDN compatible
characters are allowed. Please re-enter the label.
Start with index
Number of labels
Associated with Unit
OK
Cancel
Note
The warning message
One or more new labels are already in use. Cannot create duplicate labels
appears if duplicate labels are detected in the Add and Remove label operations.
Remove
Note
If a device is currently configured to a Label, the following message appears:
Right-hand fields in the Equipment Setup window are used to map bed labels to equipment
of various types. There are icons available for mapping bed labels to equipment that depend
on the Monitoring Type of the Unit selected and types of devices (i.e. SCCs and Switches).
There are three basic modes for mapping bed labels to equipment:
Selecting the SCC icon is used to map either hardwired or telemetry SDN based bedsides
to a bed label. SDN patient monitors on each SCC branch # and the Information Center that
receives their patient data are displayed.
Selecting the Switch icon is used to map bed labels to Mx/IntelliVue Patient Monitors in a
unit using the fixed monitoring mode where the Mx/IntelliVue is mapped to a specific port
in the switch. The switch port number, their associated Bed Label, and the Information
Center that receives their patient data is displayed.
Selecting the Monitor Label icon is used to map bed labels to Mx/IntelliVue Patient
Monitors in a unit using flexible monitoring mode, where the bedside monitors have been
Type the first number in the list, e.g. 1.
Type the number of labels that have the specified Prefix and follow in sequence from the Start
with index: number. Generally, the Number of labels will be the number of Beds, Telemetry
Equipment, or Monitors in the clinical Unit -- one number for each device
Appears for the Bed labels and is for selecting the clinical Unit where the monitoring equipment
is located. Clicking on (highlighting) the Unit name in the field selects it for assigning Bed
Labels. The up/down arrows to the right of the field permit scrolling the list up and down.
Stores the information that has been entered, which then appears in the Labels list
Closes the Add new labels window and returns to the Equipment Setup window
is for removing a label from the Label field. The procedure is as follows:
• Click on (highlight) the Label(s) in the Label column to select them for removal.
• Click Remove to remove the selected Label(s).
for Bed Labels: Cannot remove the beds selected because equipment is connected to
some or all of them.
for other Labels: Cannot remove the selected labels because beds are mapped to some
or all of them.
Software Installation & Configuration
Configuration
6-37