Table 8-3. M3150/M3155/M3151 - Test and Inspection Matrix
Test Block
Test or Inspection to Perform
Name
Performance:
Verify purchased options are enabled.
Do appropriate data (waveforms and parameters) appear?
Verify normal system operation.
Do the devices respond to user inputs?
Do audible speaker tones occur?
If a Keyboard-Video-Mouse Switch is used, has the KVM
Switch successfully passed its Performance Test?
If #C14 HL7 Export option is installed, verify successful
ping to the target HL7 client device by using hostname
If #C17 12-Lead Analysis/Export option is installed,
verify successful ping to the 12-Lead TraceMaster system
device by using hostname
If #C23 Holter Export option is installed, verify successful
Ping to Holter System by using hostname.
If #C65 Alert Data Export is installed, verify successful
ping to Alert Data Export device by using host name
Safety
Is the device connected to a 78581A/B (SCC) or telemetry
mainframe?
Is there a 2nd earth ground connected to the SCC or
telemetry mainframe
Confirm that 78581A/B or telemetry mainframe has
passed Safety Test Block (see 78581A/B or Telemetry
Mainframe Test and Inspection Matrix).
Perform 78581A/B or Telemetry Mainframe Safety
Test Block
M3170 Patient
Table 8-4 describes the test or inspection to perform for each M3170 Patient Link for each
Link
type of test specified in Table 8-2.
Note
Detailed procedures for performing the tests in the tables are given in the previous sections of
Testing Product Assurance.
Configuring the System
Expected Results
Expected answers are
"Yes".
If so, Performance test
passed.
Expected answers are
"Yes".
If so, Performance test
passed.
Expected answer is
"Yes". If so,
Performance test passed.
Expected answer is
"Yes". If so,
Performance test passed.
Expected answer is
"Yes". If so,
Performance test passed.
Expected answer is
"Yes". If so,
Performance test passed.
If answer is "No", no
Safety test required.
If answer is "Yes",
move to next step.
Expected answer is
"Yes".
If so, Safety test passed.
If answer is "Yes",
Safety test passed.
If answer is "No",
move to next step.
Expected answer is
"Yes". If so, Safety test
passed
Start-up and Configuration
What to Record
on Service
Record
P:P or P:F
where P = Pass
F = Fail
PKVM:P or
PKVM:F
where P = Pass
F = Fail
PHL7:P or PHL7:F
where P = Pass
F = Fail
P12LEAD:P or
P12LEAD:F
where P = Pass
F = Fail
PHE:P or PHE:F
where P = Pass
F = Fail
PADE:P or
PADE:F
where P = Pass
F = Fail
S:NA
where NA = Not
required
S:P or S:F
where P = Pass
F = Fail
S:P or S:F
where P = Pass
F = Fail
S:P or S:F
where P = Pass
F = Fail
8-13