Recorder Documentation Services - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

IntelliVue Information Center Application Software
Recorder
Documentation
Services
3-20
Software Description
A functional block diagram of Documentation Status Services for the Philips
Recorder is shown in Figure 3-15.
Applications Manager
Serial B
Interface
Figure 3-15 Documentation Status Services for the Philips Recorder
Communication between the Applications Manager and the Philips Recorders can be
of two types -- requests to record and requests for recorder status. Requests to record
include the name of the patient, time of request, length of recording, and waveform
data. Requests for recorder status include recorder failure, out of paper, door open,
and no recorder connected.
Process components of Documentation Status Services for the Philips Recorder do
the following:
Documentation Status Services receive recorder requests from the Applications
Manager and pass them on to the Recorder Manager. If patient data are to be sent to
the Recorder, they are obtained from the Physio Data Server.
Recorder Manager keeps track of the status and location of Recorder(s) in the
Recorder Rack and assigns recorder information to the next available Recorder.
Shared Memory provides the recorder status messages from its RAM based on
status information received from the Recorder.
Redirector adds a header message to the destination recorder.
Physio
Server
Documentation
Status
Services
Data
Recorder
Manager
Shared
Memory
Redirector
SDN
Driver
SDN
Interface

Advertisement

Table of Contents
loading

This manual is also suitable for:

M3170M3155bM3151bIntellivue seriesM3169aM3154b

Table of Contents