Philips M3150B Installation And Service Manual page 255

Information center system
Table of Contents

Advertisement

Application Server
Alert Data Export Client
SNTP Time Source
Holter
Network Printer
Wireless Access Point
Buttons to the right of the field are for adding, editing, and removing Devices from the list.
Button
Database Server
Local Database M3150 Information
Center
Database Server
Local Database M3150 Information
Center
Database Server
Local Database M3150 Information
Center
Database Server
all
all
Description
Application Server must be configured with the
Database Server or Information Center hostname
and Clinical Network IP Address. The Database
Server or Information Center must be configured
with the Application Server's hostname and IP
Address (172.31.211.0).
Device name and host name of Alert Data Export
client is required to add the client to the list of
network devices. A Port number must be defined,
and it must be unique for each DBS. The range of
port numbers is 8010-8050. The IP address of the
Alert Data Export client can be used, and if it is
entered, it can be used in place of the hostname.
SNTP Time source must be provided by the
hospital, and must not be part of the IntelliVue
Clinical Network. It is defined as a separate PC and
must be an SNTP Master Time Server that is
running the W32TIME service as defined for
Windows NT/2000/XP/ME. The host name can
either be a DNS hostname, URL, or IP Address. The
time source can reside geographically in any time
zone without any requirements for "Automatically
adjusting clock for Daylight saving changes". If the
time source is behind the hospital's firewall, the
hospital must open the appropriate port (port 123)
for the NTP protocol.
Holter device must be configured to receive
Information Center data. (PMDExport folder must
be created). Only 1 Holter device is supported per
system. The Database Server must be configured
with the 12-Lead ECG Management system
hostname.
The Printer MAC address is needed to add Network
printers. This is the only supported way to add
network printers to the Information Center system.
Must be configured using the Information Center
ConfigTool
Software Installation & Configuration