Testing Product Assurance

Note

Visual Tests

System Components

Cables

Connectors

Warning
Testing Product Assurance
Before using the Philips system clinically with patients, proper performance of the
system shall be verified. This section includes a series of Product Assurance Tests
that shall be performed after system installation and any system repair or upgrade to
verify system functionality.
These validation tests do not check all system specifications but are intended to verify
performance of the primary features of Information Center functionality. However,
successful completion of the performance tests shall yield a fully functioning system.
When performing product assurance tests, Information Centers and Clients must be
connected to the SDN or Clinical Network.
Prior to using the Philips system with patients, thoroughly inspect all system
components, cables, and connectors.
Step 1. Check all components of the Information Center system for signs of
mechanical damage.
If damage to a component is found, assess the damage to determine if repair or
replacement is required. Repair or replace the component as required before
continuing the Product Assurance Tests.
Step 2. Check all cables of the Information Center system for signs of abrasion,
wear, or other damage.
If any cable shows evidence of damage, repair or replace the cable prior to using the
Information Center system for patient monitoring.
Step 3. Check all cable connectors for signs of mechanical damage and each cable
connection for connection integrity.
If any cable connector shows signs of damage, replace the cable prior to using the
Information Center system for patient monitoring.
Step 4. Check that all cable connectors are securely fastened to the rear of each
device, including all cable ties. See Figure 5-13 Securing PC Cables.
Verify that the speaker plug is securely fastened and cannot accidentally be
pulled out. Verify the Strain Relief bracket and cable ties are attached. If a
Keyboard-Video-Mouse Switch is connected, verify that the keyboard and
mouse cables are securely fastened to the rear of the Switch and cannot
accidentally be pulled out. Refer to Chapter 5 for details.
Testing Product Assurance 8-1