Description (Event Code)
SDN
Error: 930, Plugged devices table does not match number of
devices marked as plugged in
Error: 931, Invalid rack interface number or rack slot number
Error: 935, A device other than a Morgana recorder is plugged
in
Error: 938, Can't find a device status entry that should be there
Error: 1553, SDN adapter incompatible hardware and firmware Replace SDN Interface Card. If problem
Error: 1554, SDN adapter missing hardware
Error: 1793, Adapter never responded to reset command within
time-out period. Adapter has been declared dead.
Error: 1794, Firmware data file janus.afc is not accessible.
Error: 1795, Invalid firmware found on adapter -- Loading new
firmware from local firmware file
Error: 1796, Unable to create the symbolic link for janus.afc.
Error: 1797, No SDN adapter interrupts have occurred within
time-out period - Scheduling an adapter reset.
Error: 1798, Performing driver initiated adapter reset.
Error: 1799, The SDN adapter is running.
Error: 1800, A user requested an adapter reset via the
SdnMonitor utility.
Error: 1801, The driver was not called back to perform an
adapter reset within the predetermined amount of time. The
driver and adapter have been made inactive.
Error: 1802, Timeout while waiting for the OS kernel to call
the driver back to perform an adapter reset - retrying kernel
call-back request.
Error: 1803, The previous error has repeated. To avoid filling
the event log with the same error, this error will not be logged
until a different error, or no error, is encountered.
Error: 1804, The adapter's boot strap code (DBOOT) has been
started. The driver will attempt to restart the SDN adapter.
Table 7-8. System Events
Corrective Action
Reboot Philips system. If problem persists,
replace SDN Interface Card
Connect correct or new rack to the Philips
system
Remove incorrect plug-in module
Reboot Philips system. If problem persists,
replace SDN Interface Card
persists, reinstall Philips software.
Replace SDN Interface Card.
Reboot Philips system. If problem persists,
replace SDN Interface Card
Reinstall Philips software.
No action required, normal operation when the
Philips system is started.
Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
Reboot Philips system. If problem persists,
replace SDN Interface Card
No action required, normal operation when the
Philips system is started.
No action required, normal operation when the
Philips system is started.
No action required, normal operation when the
Philips system is started.
Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
Reboot Philips system. If problem persists,
reload Philips software. If problems still persists,
reload OS and then Philips software.
No action required, normal operation when the
Philips system is started.
No action required, normal operation when the
Philips system is started.
Maintenance, Troubleshooting, and Repair