Description - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

Buttons below the field are for adding, changing, and deleting Clinical Unit names. The
maximum number of Clinical Units is 8.
Button

Description

Adds a Clinical Unit name to the list.
New
Note—M3170 Patient Link devices must be configured for Fixed monitoring
Note—TraceMaster supports Department ID: 01-99. ID 00 is "lost and found"
Note—Large network systems do not automatically check for duplicate clinical unit names or
duplicate bed labels. If a large network is being configured, verify that no duplicate names are
entered. When transferring patient data across database servers, the clinical units are not listed under
their database server. If duplicates exist, there is no way to determine which unit belongs to which
database server.
Note
At least one Clinical Unit must be entered in the Clinical Units field. A new name must be
added before an unwanted name is deleted.
Button
Description
Edit
Changes the name of a Clinical Unit in the list
Remove
Removes a name from the Clinical Units list.
SCCs
bedsides are connected. The maximum number of SCCs is 8. Names can be added, edited, and removed
• Click New to bring up the Add/Edit Unit window.
• Type in the name of the Clinical Unit to be added in the Type new name or label: field. The
name has a limit of 12 characters.
• Select the correct monitoring mode in the Monitoring field. Fixed monitoring is for systems
where dynamic movement of equipment is not expected (e.g. Patient Link). Flex is for systems
with Telemetry equipment, IntelliVue Patient Monitors, or M3/M4 bedsides that are to be
identified by their monitor label.
• Department Id is only required for systems with 12-Lead ECG Management configured. The
Department Id configured here must match the department ID configured in the ECG
Management system. This Id has a 32 character limit. The default is 00.
• Click OK and the new Clinical Unit name and its monitoring type is added to the Clinical Units
list.
• Click on (highlight) the Clinical Unit name to changed to select it.
• Click Edit to bring up an Add/Edit Unit window. (Double clicking on the Clinical Unit name
brings up this window directly).
• Type in the new name of the Clinical Unit in the Type new name or label: field.
• Change the Monitoring mode for the unit in the monitoring field if needed. Note that if
IntelliVue Patient Monitors or M3/M4 beds are already assigned, and the monitoring mode is
changed, all IntelliVue Patient Monitors and M3/M4 beds assigned to that unit must be deleted
before the monitoring mode can be changed.
• Click OK and the changes appear in the Clinical Units list.
• Click on (highlight) the Clinical Unit name to be deleted to select it.
• Click Remove and the name is deleted from the Clinical Units list. If there are beds mapped
for monitoring to this unit, those associations must be removed otherwise the Clinical Unit
cannot be removed.
lists the Communications Controllers connected to Information Centers on the Network if SDN
Software Installation & Configuration
Configuration
6-25

Advertisement

Table of Contents
loading

This manual is also suitable for:

M3170M3155bM3151bIntellivue seriesM3169aM3154b

Table of Contents