General Configuration; Description - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

Configuration
General
General Configuration is used to set several Domain-Wide clinical configuration settings
Configuration
that apply to all Information Centers and Clients on the Network. The General
Configuration window is shown in Figure 6-10.
The keyboard Tab key can be used to progress through the fields. To make a selection in fields
Note
having options, click a
Description Fields in the General Configuration window are:
Database Server Settings are set only at the Database Server and apply to all Information
Centers and Clients on the Network. They include the following:
Field
Institution Name
Language
Institution ID
Hemodynamic Pressure
Gas Pressures
Weight Units
Height Units
6-20
Software Installation & Configuration
Figure 6-10 General Configuration Window
in the circle preceding the desired option.

Description

Name of the hospital where the system resides. The Hospital Name appears in the footer of
some printed reports. This field has a 48 character limit.
Read Only. Displays the language used in the display text.
Note—The language is selected during Windows Operating System installation. To change the
language, the Operating System must be reinstalled and the desired language selected during the
installation process
Required for systems with 12-Lead ECG Management export configured. This is a 32 character
alphanumeric string that represents the Institution ID. If the 12-Lead ECG Management option is
not configured, the default value of 000 should not be changed.
Note—TraceMaster supports Institution ID: 001-998.
mmHg or kPa (millimeters mercury, kilo Pascals).
or
(millimeters mercury, kilo Pascals).
mmHg
kPa
Lbs or Kg (pounds or kilograms)
Inches or cm (centimeters)