Troubleshooting
Description
Source
(Event Code)
dataserver
Error: Unhandled
Exception Filter is
redirector
kicking in
recmgr
sdprocess
pdsstore
docstatusserver
EMSMainMenu Winsock connection failed
Mode of operation could not be read from registry. Therefore,
the mode was defaulted to monitoring mode.
The user logon flag could not be read from the registry.
Therefore, the user logon was defaulted to the
SystemManager account
Could not open the specified event to send to the Shell
application.
Could not start application because Error = XX
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Could not write value to Registry because Error = XX
Install
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Could not write value to Registry because Error = %d
OpenProcessToken failed.
AdjustTokenPrivileges enable failed.
Auto Reboot Failed
NetCfgLib
Auto Reboot failed
OpenProcessToken failed
AdjustTokenPriveleges enable failed
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Could not write value to Registry because Error = XX
Could not read value from Registry because Error = %d
Could not write to Error File
OrbPlus
Error: all
7-30
Maintenance, Troubleshooting, and Repair
Table 7-7. Application Events
Explanation
Fatal error. System will automatically
reboot to clear error state
Corrective Action
Reboot system to clear. Contact Response
Center if it happens often.
Reboot system. If this persists, reinstall OS
software.
No action required.
Reboot system
Reboot system
Reboot system. If this persists, reinstall
Philips software
Reboot system. If this persists, reinstall OS
and Philips software
Reboot system. If this persists, reinstall
Philips software
Reboot system.
Reboot system. If this persists, contact
Response Center
Contact Response Center
Contact Response Center
Reboot system or recycle power
Reboot system, If this persists, reinstall
Philips software
Reboot system. If this persists, reinstall
Philips software
Reboot system. If this persists, reinstall
Philips software
Reboot system. If this persists, reinstall OS
and Philips software
Reboot system. If this persists, reinstall
Philips software
Reboot system. If this persists, reinstall
Philips software
No action required
No action required