Configuring the System

Modem Test

The following procedure tests a factory installed modem on a PC or server and external dial-
in access.
Step 1. Click Start in the Windows Main Menu and then Programs.
Step 2. Click Accessories in the Programs menu.
Step 3. Click Hyperterminal in the Accessories menu.
Step 4. Click HyperTerminal in the Hyperterminal menu.
If the Location Information window of Figure 8-2 appears:
8-8
Start-up and Configuration
Use the ruler to measure (in millimeters) the full width (W) of a waveform in a Patient
Sector of the Main Screen display as shown below. Use the ruler measurement
techniques of Step 5 of For CRT Displays:
(W)
0
Width
Measure (in seconds) the time (t) for 10 full sweeps of the erase bar for the same
waveform as shown below. The measurement shall be to the nearest second.
(t)
Total Time of 10 Sweeps in Seconds (t)
10
To meet the sweep speed specification of 25 mm/s +10%, the measured time (t) shall be
between:
9.1 x W
25
Select a country/region
Enter an area or city code
Select whether the phone is Tone dialing or Pulse dialing
Waveform Width in Millimeters (W)
11.1 x W
<
<
t
25