Philips Carenet; Network Cabling - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

Philips CareNet

Information Centers are designed to operate on Philips CareNet network, consisting of the
Serial Distribution Network (SDN) and the Philips System Communications Controller (PCC/
SCC) or the CareNet Controller. See Figure 1-9.
The Philips CareNet is a local area communications network designed to share patient
physiological waves, parameters, and other data among bedside monitors, recorders, printers,
and other computer systems.
The PCC/SCC is the active device of the SDN, providing communications links to
instruments connected to the SDN and controlling data flow, timing, synchronization, and
distribution throughout the network. The Information Center processing unit contains an SDN
interface card for communication with the SCC and its SDN networked instruments.
Network
Network cabling between the SCC and wall boxes for system instrument connections (bedside
Cabling
monitors, central stations, etc.) can be Philips' proprietary, shielded, System Distribution
Cable (SDC) or standard LAN, Category 5, Unshielded Twisted Pair (UTP) cable.
Patient Monitors
Telemetry
Monitor
Bedside
Monitors
Information Center System
PCC/SCC
Processing Unit
Switch
Color Display
Keyboard
Figure 1-9 Typical Information Center Monitoring System Network
2-Channel Recorder
Mouse
Introducing the IntelliVue Information Center System
Philips CareNet
4 Channel Recorder
(Optional)
UPS
Printer (optional)
Speaker
1-13