Configuration
Field
MOVE (UP, DOWN)
Choose New Color
Warning
If the correct primary display is not selected, the display sweep speed will not be correct
and Philips application software will not meet its specifications.
Note
The number of waves on the Main Screen can be 3 for 4 Patient Sectors, 2 for 5-12 Patient
Sectors, and 2 for 16 Patient Sectors.
Notes
The order of the Waves or Parameters specifies their priority for display.
The priority order of Waves is fixed, but the order of Parameters can be changed.
Procedure The procedure for configuring the Display Setup window is as follows.
Step 1. Select whether the colors of the Wave and Parameters should be Standard or By
Step 2. Set the desired priority of the Physiological Parameters using one of the
Step 3. If By Parameter color selection was chosen in Step 1, set the colors of the Waves
Step 4. Click Next> to commit changes and advance to !Bed Configuration.
Note
For Information and Centers on the Network without a Database Server, the next configuration
window is Equipment Setup, for creating labels for Beds, Telemetry Equipment, and
bedside Monitors and assigning monitors to beds.
6-34
Software Installation & Configuration
Description
Use to change the priority order of the Parameters for display.
To change the order of a Parameter, there are two methods:
• Click on the Parameter to be changed to activate it for a change.
– Click the
or
UP
DOWN MOVE
• Click and hold the left mouse button on the Parameter to be changed.
– Drag the Parameter to its new location. (A gray line identifies the location of the
changing Parameter during the drag.)
– Release the left mouse button when the Parameter is positioned in its new location.
Used to change the color of Waves and Parameters.
To change the color of a Wave or Parameter:
• Click on the Wave or Parameter to be changed to activate it for a change.
• Click on the desired color for that Wave or Parameter from the 6 options provided in the
field. Changing the color of a Wave automatically changes the color of
Choose New Color
its Parameter.
• View the changed color in the
Parameter.
methods to change the order described above.
and Parameters to the desired colors using the Choose New Color box using the
method to change the color described above.
button to the right of the list to move its position in the list.
field.
Preview