Philips M3150B Installation And Service Manual page 458

Information center system
Table of Contents

Advertisement

Configuring the System
Test Block
Test or Inspection to Perform
Name
Performance:
Verify purchased options are enabled.
Observe that all M3155 and M3151 Client Device
names configured on the DBS are displayed at the
bottom of the service screen with a status of Monitoring
If #C22 Web Access option is installed, log in to the
Database Server from a PC on the hospital's intranet or
from the Database Server
Verify that stored patient data are displayed properly
If #C22 Patient Data Transfer option is installed, verify
in Add/Remove Database Servers configuration that all
DBS hostnames participating in patient data transfer are
listed.
Ping by hostname each of the Database Servers
8-16
Start-up and Configuration
Table 8-6. M3154 - Test and Inspection Matrix
Expected Results
Expected answers are
"Yes".
If so, Performance test
passed.
Expected answer is
"Yes". If so,
Performance test passed.
Expected answer is
"Yes". If so,
Performance test passed.
What to Record
on Service
Record
P:P or P:F
where P = Pass
F = Fail
PWA:P or PWA:F
where P = Pass
F = Fail
PPDT:P or PPDT:F
where P = Pass
F = Fail