Philips M3150B Installation And Service Manual page 299

Information center system
Table of Contents

Advertisement

Symptom
Possible Cause
Second NIC Card on Database
Server/Information Center is not
configured correctly.
Hospital ID/Department ID
configuration settings do not
match those of the receiving
device
Action
Verify the Second NIC Card is configured
with the appropriate settings. See "Network
Card and TCP/IP Settings" on page 4-23
Verify the configurations settings in the
Purchased Options and Support Information
Page and the Equipment Setup (clinical unit
setup) window match. See Purchased
Options and Support Information and
Equipment Setup in Chapter 6.
Maintenance, Troubleshooting, and Repair
Troubleshooting
7-13