Perform a D1 and D2 density calibration using ProLink III

1.
2.
180
c. Choose Start Calibration.
d. Wait for the calibration to complete.
e. Choose Finished.
Fill the sensor with the D2 fluid and allow the sensor to achieve thermal equilibrium.
Perform the D2 calibration.
a. Choose D2 (Water).
b. Enter the density of your D2 fluid.
c. Choose Start Calibration.
d. Wait for the calibration to complete.
e. Choose Finished.
Open the shutoff valve.
Read
page 179
if you have not already done so.
See the following figure.