Perform A D1 And D2 Density Calibration Using Prolink Iii - Emerson Micro Motion 1500 Configuration And Use Manual

Transmitters with analog outputs
Hide thumbs Also See for Micro Motion 1500:
Table of Contents

Advertisement

Restriction
For T-Series sensors, the D1 calibration must be performed on air and the D2 calibration must be
performed on water.
9.7.1

Perform a D1 and D2 density calibration using ProLink III

1.
2.
Configuration and Use Manual
If LD Optimization is enabled on your meter, disable it. To do this, choose Configure
> Manual Setup > Measurements > LD Optimization. LD Optimization is used only
with large sensors in hydrocarbon applications. In some installations, only customer
service has access to this parameter. If this is the case, contact customer service
before continuing.
The calibrations must be performed without interruption, in the order shown. Make
sure that you are prepared to complete the process without interruption.
Before performing the calibration, record your current calibration parameters. You
can do this by saving the current configuration to a file on the PC. If the calibration
fails, restore the known values.
Read the Prerequisites on
See the following figure.
page 114
if you have not already done so.
Measurement support
115