Measurement support
Run an SMV test using a web browser
1.
2.
3.
4.
Postrequisites
View the results and take any appropriate actions.
12.1.2

View SMV test results

After each SMV test, the pass/fail result is displayed automatically. Detailed results are also
available for licensed versions.
Tip
For licensed SMV versions only, when you use either the display or the web browser to view test
results, the 20 most recent results are available. If you use ProLink III to view results, you can view
results for all tests that are in the PC database.
164
Read the Prerequisites in
Choose Service Tools > Smart Meter Verification > Run Meter Verification.
Select the desired output behavior.
Option
Description
Continue Measuring During the test, all outputs will continue to report their assigned proc-
ess variables. The test will run for approximately 90 seconds.
Fix at Last Meas-
During the test, all outputs will report the last measured value of their
ured Value
assigned process variable. The test will run for approximately
140 seconds.
Fix at Fault
During the test, all outputs will go to their configured fault action. The
test will run for approximately 140 seconds.
Click Start and wait for the test to complete.
Tip
At any time during the process, you can abort the test. If the outputs were fixed, they will
return to normal behavior.
When SMV is licensed, results for the test are stored in the transmitter memory,
along with results for tests performed with any other tool. These results are also
stored in the ProLink III database. You can view and use these results in any tool-
based trending and reporting functions.
When SMV is not licensed, results from previous tests are saved in a database, but
not accessible. A Pass/Fail indication is all that is provided for the current test.
View SMV test results using the display
View SMV test results using ProLink III
View SMV test results using a web browser
Section 12.1
if you have not done so already.